CLIA Kit for Neutrophil gelatinase-associated lipocalin (NGAL)
Product Specifications
Product Name Alternative
Lipocalin-2; LCN2; p25; Lipocalin 2; Oncogene 24p3; 25 kDa alpha-2-microglobulin-related subunit of MMP-9; Siderocalin LCN2
UniProt
P80188
Applications
Chemiluminescent immunoassay for Antigen Detection.
Field of Research
Infection immunity
Assay Principle
The microplate provided in this kit has been pre-coated with an antibody specific to Neutrophil gelatinase-associated lipocalin (NGAL) . Standards or samples are then added to the appropriate microplate wells with a biotin-conjugated antibody specific to Neutrophil gelatinase-associated lipocalin (NGAL) . Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is proportional to the Neutrophil gelatinase-associated lipocalin (NGAL) level in the sample or standard.
Assay Protocol
1. Prepare all reagents, samples and standards; <br/> 2. Add 100µ L standard or sample to each well. Incubate 2 hours at 37° C; <br/> 3. Aspirate and add 100µ L prepared Detection Reagent A. Incubate 1 hour at 37° C; <br/> 4. Aspirate and wash 3 times; <br/> 5. Add 100µ L prepared Detection Reagent B. Incubate 30 minutes at 37° C; <br/> 6. Aspirate and wash 5 times; <br/> 7. Add 100µ L Substrate Solution. Incubate 10 minutes at 37° C; <br/> 8. Read RLU value immediately.<br/>
Assay Performance Time
2h, 40min
Sample Type
Serum, plasma, urine, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Detection Range
27.4-20,000pg/mL
Precision
Intra-assay Precision (Precision within an assay) : 3 samples with low, middle and high level Neutrophil gelatinase-associated lipocalin (NGAL) were tested 20 times on one plate, respectively. Inter-assay Precision (Precision between assays) : 3 samples with low, middle and high level Neutrophil gelatinase-associated lipocalin (NGAL) were tested on 3 different plates, 8 replicates in each plate. CV (%) = SD/meanX100 Intra-Assay: CV Inter-Assay: CV Inter-Assay: CV
Sensitivity
The minimum detectable dose of this kit is typically less than 10.2pg/mL
Stability
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition. To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Format
48T, 96T, 96Tx5, 96Tx10, 96Tx100
Specificity
This assay has high sensitivity and excellent specificity for detection of Neutrophil gelatinase-associated lipocalin (NGAL) . No significant cross-reactivity or interference between Neutrophil gelatinase-associated lipocalin (NGAL) and analogues was observed.
Product Name Synonym
Neutrophil gelatinase-associated lipocalin
Method
Double-antibody Sandwich
Organism Species
Homo sapiens (Human)
Available Sizes
Curated Selection
Explore Other Products